Product Manual Of High-precision Miniature Fluorescence Detector (industrial Grade) BLP-FD-10x

1. Features

  • Integrated excitation light, beam collimation, light leveling, extinction, fluorescence collection, fluorescence detection, filtering, signal shielding protection in one.
  • Extinction tail cone, blackened internal thread design, fully eliminating stray light and minimizing optical noise.
  • Built-in high-precision low-pass filter circuit, effectively filtering out signal noise.
  • On-board signal shield to effectively prevent interference such as electromagnetic radiation.
  • Built-in high-precision 16-bit analog-to-digital conversion circuit, fluorescent signal values are transmitted to the computer for real-time data analysis.
  • Independent design of analog power supply and digital power supply, effectively preventing crosstalk and improving signal-to-noise ratio.
  • Modular design, compact structure, easy to assemble, debug and maintain.

2. Applications

  • Microfluidic Technology
  • Nucleic Acid PCR Detection
  • POCT
  • Protein Analysis
  • Medical Analysis
  • Chemical Analysis
  • Food Safety Research
  • Water Quality Research
  • Pathogenic Microorganism Research
  • Handheld Laboratory

3. Schematic Diagram

4. Structure

5. Mechanical Dimensions

Notes:
(1) All dimensions in mm
(2) Weight 65g

6. Specifications(industrial Grade)

Parameters

BLP-FD-101A

BLP-FD-102A

BLP-FD-103A

BLP-FD-104A

Excitation Filter(nm)

475

525

575

625

Emission Filter(nm)

525

575

625

692

Optical Density

OD5

OD5

OD5

OD5

Dichroic Mirrors
Long-pass(nm)

≥490

≥550

≥580

≥650

Spot Diameter

6.5mm

6.5mm

6.5mm

6.5mm

Input Voltage

+5V or USB(PC)

+5V or USB(PC)

+5V or USB(PC)

+5V or USB(PC)

LED Current

10 – 650mA

10 – 650mA

10 – 650mA

10 – 650mA

Analog Signal Output

0 – 3.3V

0 – 3.3V

0-3.3V

0 – 3.3V

Fluorescent Dye

FAM/SYBR Green

HEX/VIC

ROX

CY5

Notes:
LED Light Sources Must Use A Constant Current Driver

7. Read the Signal Value

8.Model Naming Rules